|
Once selected the settings, the threshold should be selected.
Figure 3.2 shows the number of counts as a function of the threshold value in the ideal case of monoenergetix X-rays of energy
=10 keV.
For thresholds larger than the X-ray energy the detector should always count 0 and for lower thresholds it should always count all the photons. However the curve is smoothed around
because of the electronic noise (ENC) and is not perfectly flat for lower energies because the photons absorbed in the region between two strips distribute their energy between them and it is not flully collected by a single channel (charge sharing).
In order to count once al X-rays the threshold should be set at half of the X-ray energy
: if the threshold would be higher some photons would not be counted, leading to a loss of efficiency, while if it would be lower some photons would be counted twice leading to a loss of spatial resolution.
Since the detector threshold can't be precisely set at the same value for all channels but there will always be some spread of the order of 200 eV (threshold dispersion) there will always be some fluctuations on the number of counts between channels, which however should be corrected by the flat field correction.
The choice of the threshold should also depend from considerations regarding the emission of fluorescent radiation from the sample.
Figure 3.3 shows how the curve of the counts would look like for monochromatic X-rays of energy
in presence of radiation of energy
emitted by the sample. The curve would show a second step at
.
Since the fluorecence emission is not present in the flat field data, the difference of counts between the channels due to the fluorescent radiation cannot be corrected and the threshold
should be set at an energy larger than
. This also helps to cut down the background.
The difference of counts between the channels will be particularly large if the threshold is set in some ``steep'' part of the curve i.e. close to
or to
(but in this case it would be corrected by the flat field, at cost of loss of efficiency).
Because of the presence of the electronic noise,
should be at least 3 keV larger than
.
Here is a short list of rules to select the appropriate working threshold in order of importance (and eventually modify the X-ray energy):
|