In order to convert from strip number to 2-angle, an accurate angular calibration of the detector must be performed (for details see the paper Bergamaschi, A. et al. (2010). J. Synchrotron Rad. 17, 653-668).
For this purpose, a series of patterns of a powder standard with symmetric peaks (e.g. silicon) must acquired while shifting the detector by an angular step of the order of about 2% of the module size. During the measurement, a strong intensity peak (e.g. Si(111)) should pass through the field of view of every module such that it can be used as a reference angular position to perform the calibration of the modules position.
In a first step, the peak is fitted with a Gaussian in order to determine its position in channel number for each of the acquired patterns.
In a second step, for each module , the encoder position is fitted as a function of the peak position according to:
The same function of equation 1, with the parameters obtained from the calibration, is used in order to convert from channel number to 2-angle.
The parallax at the borders of the modules due to the thickness of the silicon sensor is a function of the X-ray energy (higher energy X-rays are absorbed deeper inside the sensor) and is of the order of 0.2 mdeg at 12 keV and 0.5 mdeg at 30 keV.
The differences in pixel size due to the different portion of solid angle covered by the strips on the border of the modules and the higher efficiency due to the longer path of the X-rays in the sensor are removed by the flat field correction. This also normalizes additional differences in pixel size between channels which are also present because of mismatches in the strip sensor fabrication and in fluctuations of the channels threshold level.
Patterns acquired at different detector positions are generally merged together in order to fill the gaps between the modules and correct possibly bad functioning channels. In this procedure the data from different positions which are closer than 4 mdeg (the average pixel size) are averaged and the new position is set to the mean of the positions of the original points.
The position and width of the peaks results from a fit over several detector channels. Geometrical distortions might disturb this determination mainly because of errors in the angular calibration, fluctuations in the encoder position, variations between channels and parallax effects.
The resolution in locating the peak center and determining its width and integrated intensity has been estimated by acquiring several patterns of a LaB sample in a 300 m capillary with the detector shifted in 5 mdeg steps between 30.4 and 36.5 degrees. The 16 peaks acquired have been fitted with a Gaussian function plus background and the fluctuations on the fitted parameters have been calculated. The resulting average resolutions are 0.630.06 mdeg for the peak center and 0.220.05 mdeg for the peak Full-Width at Half-Maximum (FWHM) for an average peak FWHM of 27.02.5 mdeg.
These results show that the angular calibration allows a resolution in determining the peaks position and width which is appropriate for structural determination.