Getting Started =============== Detector Servers include: * Control server [default port: 1952] * Almost all client communication. * Stop server [default port: 1953] * Client requests for detector status, stop acquisition, temperature, advanced read/write registers. When using a blocking acquire command (sls_detector_acquire or Detector::acquire), the control server is blocked until end of acquisition. However, stop server commands could be used in parallel. Location --------- slsDetectorPackage/serverBin/ folder in every release. .. _Detector Server Arguments: Arguments --------- .. code-block:: bash Possible arguments are: -v, --version : Software version -p, --port : TCP communication port with client. -g, --nomodule : [Mythen3][Gotthard2] Generic or No Module mode. Skips detector type checks. -f, --phaseshift : [Gotthard] only. Sets phase shift. -d, --devel : Developer mode. Skips firmware checks. -u, --update : Update mode. Skips firmware checks and initial detector setup. -i, --ignore-config : [Eiger][Jungfrau][Gotthard][Gotthard2][Moench] Ignore config file. -m, --master : [Eiger][Mythen3][Gotthard][Gotthard2] Set Master to 0 or 1. Precedence over config file. Only for virtual servers except Eiger. -t, --top : [Eiger] Set Top to 0 or 1. Precedence over config file. -s, --stopserver : Stop server. Do not use as it is created by control server .. _Automatic start servers: Automatic start ------------------ One can start the on-board detector server automatically upon powering on the board. #. Create a soft link to the binary on board : .. code-block:: bash ln -sf someDetectorServervx.x.x someDetectorServer #. Do the following depending on the detector type : Eiger .. code-block:: bash # create script in rc5.d on the board vi /etc/rc5.d/S50board_com.sh # enter the following (edit server name) #! /bin/sh /home/root/executables/eigerDetectorServer &> /dev/null & exit 0 Jungfrau | Moench | CTB | Gotthard I .. code-block:: bash # Edit inittab on board vi /etc/inittab # enter the following line ttyS0::respawn:/./xxxDetectorServer Gotthard II | Mythen III .. code-block:: bash # create script in init.d on board vi /etc/init.d/S99detServer.sh # enter the following (edit server name) #! /bin/sh cd /root >> /dev/null /root/xxxDetectorServer >> /dev/null & #. Sync, reboot and verify : .. code-block:: bash sync # physically reboot for Gotthard II or Mythen III reboot # verify ps -ef | grep xxxDetectorServer